Volta Electra > Anti-Emulsification Performance Tester
Introduction
This instrument complies with GB / T7305, ASTM D 1401 for the determination of water separabilityof petroleumoil and synthetic liquid. It is applied to test samples whose kinematic viscosity is 30~100mm2/s at 40℃andthetest temperature was 54 ± 0.1 ℃. It also can be used for oil whose viscosity is above 100mmm2/sandthetesttemperature is 82±0.1 ℃.
Technical Specification
1. Input voltage: AC220V±20V, 50Hz 2. Input power: 1.5kW 3. Temperature adjustment range: room temperature ~90°C 4. Temperature control accuracy: 54±0.1°C; 82±0.1°C 5. Bath stirring motor: 15W 6. Sample stirring motor: 1500±15rpm 7. Automatic timing: 5 minutes 8. Structure: 2-hole rotary table, slide bar stirrer
Working Principle
The instrument uses a digital temperature controller with a thermal resistance sensor to control thetemperature.By stirring motor, the bath reaches the temperature required for the test (54±0.1°C; 82±0.1°C), andisequippedwith an automatic timing device to achieve high temperature control accuracy. Intuitive display, goodreliability,saving test time.